Post helium integrity test properties and performance of 3D storage and mixer biocontainers made with Allegro™ film

Oct 27, 2026

Introduction

This technical report contains data applicable to Allegro™ 3D storage biocontainers, LevMixer™ biocontainers made with Allegro film, and Magnetic mixer biocontainers made with Allegro film, which have been tested by a helium integrity test (using HIT™ technology) to increase assurance of integrity. The purpose of this report is to:

  1. Demonstrate comparable performance of 3D biocontainers made with Allegro film produced with and without a helium integrity test, through test data and manufacturer risk assessment.
  2. Confirm that 3D mixer biocontainers made with Allegro film incorporating protective top hat covers meet all relevant product performance specifications through test data and manufacturer risk assessment.

This report should be used to assist you in making risk assessments for processes incorporating helium integrity tested 3D storage and mixer biocontainers made with Allegro film.

Allegro biocontainers

Allegro biocontainers have been developed to replace stainless steel tanks and carboys in process applications, to reduce the risk of cross-contamination and eliminate cleaning and cleaning validation.

The Allegro biocontainer is made of a low density polyethylene (LDPE) film that meets the rigorous quality requirements of the biotechnology and pharmaceutical industry. The Allegro biocontainer includes unique design features that significantly improve the ease of use and robustness of single-use systems while enabling high product recovery. The ports of Allegro 3D biocontainers have been configured so that maximum choice is offered for inlet and outlet connections. There are two inlet ports: one out of the two inlet ports can be used either for the introduction of a sensor or as an additional filling line. Both the inlet ports and the outlet ports can be connected to tubing of ⅜, ½, ¾, and 1 inch nominal inner diameter. A ¼ inch sampling port is available for all options which could be kept open or closed depending on the specific application.

Helium integrity test (using HIT technology)

We constantly strive to meet your needs as a biopharma drug manufacturer. Although Quality by Design (QbD) principles, junction tests, quality control tests, and validated processes are used for all our manufacturing procedures, when extra risk mitigation is required, we can meet that need. Our high sensitivity helium integrity test system (using HIT technology), is implemented at the our Medemblik (Netherlands) single-use systems (SUS) manufacturing site. The HIT system was specifically designed to set the highest integrity test standard for biocontainers and the joints of the transfer tubes to the biocontainer, capable of detecting defects as small as 2 μm. For customers needing a very high level of assurance for integrity of their SUS, the helium integrity test can be performed upon request for single-use assemblies. The HIT system is designed to provide the highest level of sensitivity for SUS incorporated in process steps that incorporate SUS, especially important for the most critical step of bulk drug substance or drug product storage.

We offer the helium integrity test for Allegro single-use systems only upon specific customer requests, and after verification that their actual design is compatible with the HIT system and test procedures. The test occurs at our site during the manufacturing of the Allegro SUS.

Protective top hat for 3D mixer biocontainers made with Allegro film

During the HIT process there could be a risk that during installation or operation of the HIT method, the impeller could be in contact with the biocontainer film and cause damage if excess force was applied between these two components. To eliminate this risk, a rigid plastic protective top hat molded part was designed to be welded into the top of the mixer biocontainer to protect the biocontainer film from the impeller when the biocontainer is deflated and throughout the HIT process within the chamber.

The protective top hat can be seen in Figure 1 and surrounds the impeller when the biocontainer is deflated during storage and transportation.

Illustration of an inflated 3D mixer biocontainer with protective top hat.

Fig 1. Exterior surface of an inflated 3D mixer biocontainer made with Allegro film, with the protective top hat circled.

Summary of conclusions

Risk assessments have been conducted and conclude that neither the addition of a protective top hat nor the application of the HIT process impact the critical performance of the 3D storage and mixer biocontainers made with Allegro film. The following elements have been considered:

  • Mechanical strength.
  • Resistance to gamma sterilization.
  • Robustness during filling, drainage, transport, stacking (storage only), and mixing (mixer only).
  • Functional mixer operation (mixer only).
  • Connection strength.
  • Shelf life.
  • Extractables/leachables.
  • Biological safety.
  • Physicochemical suitability.

For verification of the consistent performance of 3D biocontainers made with Allegro film and processed with the HIT system, including mixer biocontainers made with Allegro film and the top hat molded part, the following tests were performed on biocontainers manufactured with and without the HIT process:

  1. Quantitative tensile testing of biocontainer film and welds.
  2. Endotoxin content testing.
  3. Particulate cleanliness analysis.

  4. Test 1 was carried out using storage biocontainers to test the standard film and welding, Tests 2 and 3 were carried out using mixer biocontainers with top hats to confirm their properties and performance and the results are considered to be applicable to 3D storage biocontainers which have fewer welded parts.

    For confirmation of the performance of 3D mixer biocontainers made from Allegro film and with the top hat molded part, the following additional tests were carried out using mixer biocontainers with the top hat molded part:

  5. Tensile testing of the weld between the top hat molded part and Allegro film.
  6. Package product transportation testing of mixer biocontainers with top hat molded parts.
  7. Deployment testing of mixer biocontainers with top hat molded parts.

All verification and confirmation tests passed for all biocontainers tested, made with or without the HIT process. No significant impact of the HIT process was observed for all tests. Based on the risk assessments, complemented by the above verification tests, it is confirmed that:

  • The HIT process step has no impact on properties or performance of Allegro 3D storage and mixer biocontainers.
  • 3D mixer biocontainers made with Allegro film and with the additional protective top hat meet all relevant internal specifications for existing mixer biocontainers.

Protective top hat risk assessment

The addition of a rigid plastic protective top hat is a risk-reduction strategy to eliminate the risk of damage to the biocontainer film from mixer impellers. The risk of this impacting the biocontainer performance properties is minimal because the specific plastic resin used is identical to that of the impeller, and it uses the same component manufacturing processes. In addition, the part is welded into the biocontainer using the same processes as the impeller seat. With no new materials of construction or production techniques required, the properties and performance of 3D mixer biocontainers made with Allegro film and with a top hat molded part are expected to be the same as the standard mixer biocontainers.

Table 1 shows a qualitative summary of the risk assessment for existing biocontainer properties and details the additional testing selected to demonstrate that the key claims are not affected by the introduction of the top hat.

Table 1. Risk assessment summary for the impact of the protective top hat on properties of 3D mixer biocontainers made with Allegro film

Potential failure mode Biocontainer properties impacted Existing control measures Additional testing required
Mechanical strength reduced due to incorporation of a new molded part. Mechanical strength, resistance to gamma irradiation, robustness, shelf life. Existing and effective materials of construction and welding methods. Confirm top hat seal tensile strength according to standard procedures.
Damage to the biocontainer film during packaging and transportation. Mechanical strength, robustness. Top hat design, packaging design, packaging and handling standard procedures. Confirm integrity and cosmetic condition of biocontainer manufactured with a top hat after transportation.
Endotoxin content in the biocontainer increases due to contamination from the new part. Physicochemical suitability. Existing materials of construction and component manufacturing methods. Confirm endotoxin content for mixer biocontainers manufactured with a top hat.
Additional trace impurities are added from the new part. Extractables/leachables, physicochemical suitability. Existing materials of construction and component manufacturing methods. Confirm particulate counts for mixer biocontainers manufactured with a top hat.
Large additional molded part prevents biocontainer from deploying properly during inflation. Suitability for purpose. Instructions for use. Verify proper deployment of mixer biocontainers manufactured with a top hat when inflated with air and water.

Impact of HIT process risk assessment

Exposure to helium gas under a limited and controlled pressure differential, as per the HIT process, poses a minimal increase in risk of a change in biocontainer performance. The inert nature of the helium gas ensures that a chemical change to materials does not occur, and that the potential remaining traces of helium in the biocontainer will not present a risk of interaction with the pharmaceutical product. The use of pure, filtered gas in the manufacturing cleanroom minimizes any risk of impacting physicochemical suitability. The low differential pressures controlled during the cycle ensure excess stress on the biocontainers is avoided. Also, handling procedures and safeguards are in place as per the full manufacturing process.

Table 2 shows a qualitative summary of the risk assessment for existing biocontainer properties and details the additional testing selected to demonstrate that the key claims are not affected by the HIT process.

Table 2. Risk assessment summary for the impact of the manufacturing HIT process step on properties of 3D storage and mixer biocontainer made with Allegro film

Potential failure mode Biocontainer properties impacted Existing control measures Additional testing required
Damage to the film due to excess force from the mixer biocontainer impeller during HIT process installation and vacuum cycles during operation. Mechanical strength, resistance to gamma irradiation, robustness. Manufacture all mixer biocontainers with a protective top hat. See Table 1.
Mechanical strength of biocontainer film and film welds reduced due to exposure to a HIT cycle or damage during the HIT process. Mechanical strength, resistance to gamma irradiation, robustness, shelf life. Controlled test with low pressure differential, detailed standard operating procedures (SOPs) for biocontainer handling. Compare biocontainer film tensile strength for biocontainers manufactured with and without the HIT process.
Mechanical strength of connections reduced due to exposure to a HIT cycle or damage during the HIT process. Mechanical strength, resistance to gamma irradiation, robustness, connection strength, shelf life. Controlled test with low pressure differential, detailed SOPs for biocontainer handling. No additional tests required.
Change in chemical properties of the biocontainer plastic parts due to reaction or degradation during exposure to a HIT cycle. Shelf life, extractables/leachables, biological safety, physicochemical suitability. Pure, filtered, inert gas. No additional tests required.
Endotoxin content in the biocontainer increases due to contamination during the HIT process. Physicochemical suitability. Operation in cleanroom; pure, inert, filtered gas. Compare endotoxin content for biocontainers manufactured with and without the HIT process.
Additional trace impurities are added to the biocontainer during exposure to a HIT cycle. Extractables/leachables, physicochemical suitability. Operation in cleanroom; pure, inert, filtered gas. Compare particulate counts for biocontainers manufactured with and without the HIT process.

Quantitative tensile testing of biocontainer film and welds (test 1)

Introduction

Tensile testing of films and welds is a standard approach to verify absence of impact of a processing step on a biocontainer's physical performance. Based on risk assessment, the chance of significantly reducing the tensile strength of the biocontainers is minimal due to the use of a controlled test with low pressure differential and detailed SOPs for biocontainer handling. A significantly lower tensile strength could impact claims such as mechanical strength, resistance to gamma irradiation, resistance to freezing, and shelf life. A tensile strength study was carried out in order to verify the risk assessment. The tensile strength results for the storage biocontainer are considered as representative for mixer biocontainers because of the same film material and welding methods.

Summary of methods

Two samples of a 50 L Allegro 3D biocontainer were manufactured for these tests. One biocontainer was processed through a HIT cycle and successfully passed. The other biocontainer was processed as standard. Both biocontainers were gamma irradiated at a normal routine irradiation dose of 25 to 50 kGy.

The tensile strength of the welded seal of each biocontainer was checked using a tensometer. The sample strip was 25 mm wide and the tensile force was applied at an elongation rate of 200 mm/min. During this test, the force at which the film broke was recorded. The strength of the weld at break must exceed 70 Newtons (N) to meet our internal specification.

Results

For each biocontainer, 32 individual test samples were cut from various points around the welded seal (16 from the back of the biocontainer and 16 from the front). The sample tensile strengths were measured. The minimum, maximum, and average result for each biocontainer is shown in Table 3.

Among all the samples that were tested, the polymer film material broke before any breakage of the welded seal. This is as expected, and therefore any variation in the data relates to the film tensile strength. All of the values found during tensile testing were > 70 N, confirming that every weld exceeded the specification.

A t-test (one-tailed, α = 0.05) was carried out comparing the individual tensile strength results (n = 32 for each biocontainer) and confirmed that the tensile strength of the biocontainer film is not statistically significantly lower after the HIT process. A one-tailed t-test was selected because this gives a greater chance of detecting a difference and only a reduction in tensile strength was considered to negatively impact the biocontainer physical properties.

Table 3. Tensile test on film-to-film welding of biocontainers processed with and without the HIT process1

Biocontainer test number HIT process Minimum tensile strength (N) Maximum tensile strength (N) Average tensile strength (N)
1 Yes 93.2 125.2 111.1
2 No 89.6 118.7 107.3

1 Force in Newtons to break a sample strip of 25 mm width.

Conclusion

The polymer film and welds used to manufacture HIT processed Allegro 3D biocontainers have a tensile strength exceeding the acceptance criteria. No significant differences in tensile strength at break have been observed between the HIT sample and the control sample. Absence of any impact of the HIT processing step is therefore confirmed for tensile testing characteristics of the Allegro 3D biocontainer film and its seam welds in storage and mixer biocontainers.

Tensile testing of top hat molded part welding on Allegro film (test 4)

Introduction

The addition of the top hat molded part is essential to the application of the HIT process for mixer biocontainers. Verification of the weld strength of this new part was carried out in this section without the application of the HIT process, the impact of which is covered in the previous section. Tensile testing of the welding between molded parts and the biocontainer is a standard approach to verify the absence of an impact of adding that part on biocontainer physical performance. Based on risk assessment, a tensile strength study of the weld between the biocontainer film and the top hat molded part is required. A low tensile strength could impact claims such as mechanical strength, resistance to gamma irradiation, molded part seal strength and shelf life.

Summary of methods

Thirty samples of a top hat molded part welded onto Allegro film were produced for these tests. Ten samples each were generated using low, nominal and high weld settings. Five of the samples from each weld setting were tested before gamma irradiation and the other five samples from each weld setting were tested after gamma irradiation at a normal routine irradiation dose of 25 to 50 kGy.

The tensile strength of the welded seal of each top hat molded part was checked by manually applying force according to clearly defined internal quality procedures. The weld must not break under the manual force applied in order to meet our internal specification.

Results

For each top hat molded part, four individual test samples were cut from positions equally spaced around the circumference of the part. All sample sections from all top hat molded parts tested passed the tensile strength test according to the internal criteria Q-I-PR03-0004 as detailed in Table 4.

Table 4. Results of tensile testing study on top hat molded parts welded on Allegro film

    Manual tensile test result according to Q-I-PR03-0004
Weld settings Gamma irradiation Pass Fail
Low None 20 0
Low 25 to 50 kGy 20 0
Medium None 20 0
Medium 25 to 50 kGy 20 0
High None 20 0
High 25 to 50 kGy 20 0

Conclusion

The top hat welds used to manufacture 3D mixer biocontainers with Allegro film meet the internal specifications. The mechanical strength of mixer biocontainers is not impacted by the addition of a top hat molded part.

Endotoxin content testing (test 2)

Introduction

Endotoxin testing is a critical part of physicochemical suitability tests for biocontainers and is a standard approach to verify the absence of an impact of a processing step on a biocontainer's cleanliness. Based on risk assessment, the chance of endotoxin contamination is minimal due to operation of the HIT process in cleanroom and the use of a pure, inert, filtered gas. A study was completed to further verify the endotoxin level in biocontainers manufactured with and without the HIT process. By carrying out this testing with mixer biocontainers with top hat molded parts, the endotoxin content of these biocontainers is also assessed. The mixer biocontainer results are considered representative of storage biocontainers due to the shared manufacturing materials, methods, and locations.

Summary of methods

Samples of biocontainers made with Allegro film were subjected to validated endotoxin determinations in biocontainers of volumes in accordance with USP <85> and the European Pharmacopoeia (2.6.14, current edition). The endotoxin tests were validated using chromogenic endpoint techniques with a 0.005 EU/mL sensitivity. The 50 L biocontainers were extracted with 500 mL of endotoxin free water and 15 mL of the extraction fluid was tested for endotoxin content. The data reported is the concentration of endotoxin in this extract solution. Our internal acceptance criteria for this test is ≤ 0.25 EU/mL in the extraction solution.

Results

Ten samples of a 50 L LevMixer 3D biocontainers made with Allegro film were manufactured for these tests. Five biocontainers were processed through a HIT cycle and successfully passed. The remaining biocontainers were processed as standard. All biocontainers were gamma irradiated at a normal routine irradiation dose of 25 to 50 kGy. The endotoxin levels for the biocontainers are summarized in Table 5.

Table 5. Data on endotoxin concentration in the extract solution measured from 50 L LevMixer 3D biocontainers made with Allegro film, manufactured with and without the HIT process

  Endotoxin level (EU/mL)
Biocontainer test number With HIT process Without HIT process
1 < 0.005 < 0.005
2 0.018 0.023
3 < 0.005 0.006
4 < 0.050* 0.010
5 < 0.005 < 0.005
Maximum value reported < 0.050 0.023

* Value obtained after repeat, according to procedure, as first reading was under limit of detection.

All measurements confirmed the presence of endotoxins, generally below the limit of sensitivity of the method. For 50 L biocontainers, the maximum endotoxin value of biocontainers manufactured with the HIT process (a total of five) was < 0.050 EU/mL in the 500 mL extraction volume solution. Comparable control biocontainers manufactured without the HIT process (a total of five) gave a maximum endotoxin value of 0.023 EU/mL in the 500 mL extract solution. Small values < 0.050 EU/mL observed are not uncommon. All endotoxin concentrations in the extract solution were ≤ 0.25 EU/mL, confirming that every biocontainer was clearly below the specification.

Conclusions

The endotoxin levels in Allegro 3D biocontainers manufactured with the HIT process are consistently below the acceptance criteria and comparable controls manufactured without the HIT process. Absence of any impact of the HIT process step is therefore confirmed for endotoxin cleanliness of 3D storage and mixer biocontainers made with Allegro film.

3D mixer biocontainers made with Allegro film and incorporating top hat molded parts meet the endotoxin content specifications.

Particulate cleanliness analysis (test 3)

Introduction

Particulate cleanliness is another critical part of physicochemical suitability tests for biocontainers and is a standard approach to verify the absence of an impact of a processing step on a biocontainer's cleanliness. Based on risk assessment, the chance of particle contamination is minimal due to operation of the HIT process in cleanroom and the use of a pure, inert, filtered gas. A study was completed to further verify particulate levels in biocontainers manufactured with and without the HIT process. The mixer biocontainer results are considered representative of storage biocontainers due to the shared manufacturing materials, methods, and locations.

Summary of methods

Subvisible particles content in the containers has been verified, extracting the biocontainers with a 500 mL volume solution in three consecutive rinses and applying a microscopic particle count (USP <788> test method 2) test method on each 500 mL rinse solution. The first rinse for each biocontainer was tested after removal of a 15 mL sample for endotoxin testing, evaluating the particle count in the remaining 485 mL. Our acceptance criteria for this subvisible particle testing is ≤ 12 particles/mL for particle size ≥ 10 μm and ≤ 2 particles/mL for particle size ≥ 25 μm.

Results

Ten samples of a 50 L LevMixer 3D biocontainer made with Allegro film were manufactured for these tests. Five biocontainers were processed through a HIT cycle and successfully passed. The remaining biocontainers were processed as standard. All biocontainers were gamma irradiated at a normal routine irradiation dose of 25 to 50 kGy.

The particulate testing results are summarized in Table 6 and Table 7. All data are reported in particles/mL, calculated by dividing the total particles count (sum of the three rinse solutions) by the nominal volume of the biocontainer and multiplying it by a correction factor taking into account the extraction efficiency and the proportion of sample removed from the first rinse for endotoxin testing.

Table 6. Data on > 10 μm particulate concentration in the extract solution measured from 50 L LevMixer 3D biocontainers made with Allegro film, manufactured with and without the HIT process

  Particulate concentration (count/mL)
Biocontainer test number With HIT process Without HIT process
1 0.04 0.05
2 0.05 0.10
3 0.04 0.07
4 0.07 0.05
5 0.05 0.06
Mean 0.05 0.07

Table 7. Data on > 25 μm particulate concentration of the extract solution measured from 50 L LevMixer 3D biocontainers made with Allegro film, manufactured with and without the HIT process

  Particulate concentration (count/mL)
Biocontainer test number With HIT process Without HIT process
1 0.02 0.02
2 0.02 0.04
3 0.02 0.03
4 0.03 0.02
5 0.02 0.03
Mean 0.02 0.03

All measurements confirmed a very low level of particles in the biocontainers, well below the acceptance limit. No significant difference has been observed between biocontainers manufactured with the HIT process and control biocontainers manufactured without the HIT process. Sample sizes are too small to generate a reliable t-test result, but the data for biocontainers manufactured with the HIT process are marginally lower than the control data. Any variability at this low level does not impact the overall cleanliness of the single-use system.

Conclusions

The particulate levels in 3D biocontainers made with Allegro film manufactured with the HIT process are consistently very low in comparison to the acceptance criteria and are comparable to controls biocontainers manufactured without the HIT process. Absence of any impact of the HIT process step is therefore confirmed for particulate cleanliness of the 3D storage and mixer biocontainers made with Allegro film.

3D mixer biocontainers made with Allegro film incorporating a top hat molded part meet the particulate cleanliness specifications.

Packaged product transportation testing of mixer biocontainers with top hat molded parts (test 5)

Introduction

Standard transportation test procedures from the International Safe Transit Association (ISTA) allow simulation of worldwide transportation to evaluate whether an additional new part could cause damage during shipping. Based on risk assessment, the design of the protective top hat for prevention of damage from the impeller and the standard procedures for packaging of biocontainers deliver minimal risk of damage during transportation. A transportation study on final packaged product was carried out to verify biocontainer integrity after shipping. The additional molded parts within mixer biocontainer SUS are considered as a worst case for transportation damage and the results are therefore representative for storage biocontainers.

Summary of methods

A total of 24 LevMixer 3D biocontainers made with Allegro film of varying sizes were manufactured and packaged according to our standard procedures. The biocontainers were tested for integrity during manufacturing with the HIT process and all biocontainers passed. The final packaged products were stacked onto two separate pallets in the layout specified in Figure 2.

Schematics of pallet layout for packaged biocontainers for transportation testing.

Fig 2. Pallet layout of packaged LevMixer 3D biocontainers for transportation testing.

The pallets were exposed to transport simulation testing according to (ISTA) procedure 3E: Similar packaged products in unitized loads for truckload shipment. Pallet 1 was preconditioned at 60°C ± 2°C and 30% ± 5% relative humidity for a minimum of 72 h before transport simulation. Pallet 2 was preconditioned at -29°C ± 2°C for a minimum of 72 h before transport simulation. These are the widest range of temperatures considered by ISTA 3E testing.

After preconditioning the following sequence of tests were carried out according to ISTA 3E:

  1. Horizontal impact shock at 1.2 m/s on all four sides.
  2. Rotational edge drop from a height of 230 mm onto two connecting perpendicular edges.
  3. Random vibration testing for 240 min at a root mean square acceleration of 0.54 Grms to simulate worldwide transport using a steel sprung truck.
  4. Rotational edge drop from a height of 230 mm onto two connecting perpendicular edges.

Following ISTA 3E transportation tests, all biocontainers were unpackaged and checked for cosmetic defects according to standard manufacturing quality control procedures. The worst-case location biocontainers from the top and bottom of the pallets were retested on the HIT system using the standard manufacturing test procedures.

Results

All biocontainers passed internal visual cosmetic defect evaluation and met the standards of cosmetic quality required to release manufactured products as detailed in Table 8.

Biocontainers from the top (boxes 4, 5, 6, and 7) and the bottom (boxes 1 and 10) of the pallets are considered as worst-case positions according to ISTA 3E. These biocontainers were retested for integrity using the HIT process. All biocontainers subjected to a valid helium integrity test passed as shown in Table 8.

One mixer biocontainer was not successfully tested, but exposed to a helium cycle. According to the standard manufacturing procedure for the HIT process under which these tests were run a biocontainer cannot be retested after exposure to a helium test cycle. The other three biocontainers from the same box passed the helium integrity test.

Table 8. Results of cosmetic testing and helium integrity testing of biocontainers after transport simulation

        Cosmetic test result
Preconditioning Box number Biocontainer capacity Biocontainer number Outer bag Inner bag Biocontainer HIT result
60°C/30% 1 100 L 1 Pass Pass Pass Pass
60°C/30% 1 100 L 2 Pass Pass Pass Pass
60°C/30% 1 200 L 3 Pass Pass Pass Pass
60°C/30% 1 200 L 4 Pass Pass Pass *
60°C/30% 2 100 L 1 Pass Pass Pass -
60°C/30% 3 100 L 1 Pass Pass Pass -
60°C/30% 4 50 L 1 Pass Pass Pass Pass
60°C/30% 4 50 L 2 Pass Pass Pass Pass
60°C/30% 4 50 L 3 Pass Pass Pass Pass
60°C/30% 5 50 L 1 Pass Pass Pass Pass
60°C/30% 5 50 L 2 Pass Pass Pass Pass
60°C/30% 5 50 L 3 Pass Pass Pass Pass
-29°C 6 50 L 1 Pass Pass Pass Pass
-29°C 6 50 L 2 Pass Pass Pass Pass
-29°C 6 50 L 3 Pass Pass Pass Pass
-29°C 7 50 L 1 Pass Pass Pass Pass
-29°C 7 50 L 2 Pass Pass Pass Pass
-29°C 7 50 L 3 Pass Pass Pass Pass
-29°C 8 100 L 1 Pass Pass Pass -
-29°C 8 100 L 2 Pass Pass Pass -
-29°C 9 100 L 1 Pass Pass Pass -
-29°C 10 200 L 1 Pass Pass Pass Pass
-29°C 10 200 L 2 Pass Pass Pass Pass
-29°C 10 200 L 3 Pass Pass Pass Pass

* The biocontainer integrity was not successfully tested. Retest not possible according to procedure.

Conclusions

After transportation simulation testing according to ISTA procedure 3E, all biocontainers displayed no observed sign of cosmetic defects. All biocontainers successfully tested for integrity by the HIT process passed both during production and after transportation simulation. Transportation of packaged 3D mixer biocontainers made from Allegro film and with the top hat molded part does not cause damage to the biocontainer.

Deployment testing of mixer biocontainers with a top hat molded part (test 6)

Introduction

Risk assessment indicates that the changes to the way the uninflated mixer biocontainer is folded with the protective top hat molded part incorporated could increase the chance of film catching on the impeller during inflation. This was considered during design of the top hat fitting and it is not expected to impact the proper operation according to the instructions for use. Testing was carried out to verify proper inflation during installation.

Summary of methods

Two LevMixer 3D biocontainers made with Allegro film and with top hat molded parts and two Magnetic Mixer 3D biocontainers made with Allegro film and with top hat molded parts were manufactured and gamma irradiated at a normal routine irradiation dose of 25 to 50 kGy. One biocontainer of each mixer type was inflated with air and one biocontainer of each mixer type was inflated with water according to standard instruction of use.

Results

The observations from the deployment testing are summarized in Table 9.

Table 9. Results of deployment testing of 3D mixer biocontainers made with Allegro film and with top hat molded parts

Mixer biocontainer type Inflation fluid Manual intervention required Cosmetic damage
LevMixer biocontainer Air None None
LevMixer biocontainer Water None None
Magnetic Mixer biocontainer Air None None
Magnetic Mixer biocontainer Water None None

Conclusions

LevMixer and Magnetic Mixer 3D biocontainers made with Allegro film and with protective top hat molded parts inflate normally during standard operation with either air or water. No impact of the protective top hat molded part on the successful deployment of mixer biocontainers was confirmed.


CY61211