作者:Yoshiyuki Tanaka, Hiromu Toyoda et al.
发表年份:2016年
发表期刊: Analytical Biochemistry
作者单位:泽井制药株式会社生物研究部
文章摘要
We developed a technique for the measurement of surface plasmon resonance (SPR) to detect interactions of anti-tumor necrosis factor (TNF) agents with transmembrane TNF-α (mTNF-α) on living whole cells. The injection of a suspension of mTNF-α expressing Jurkat cells, used as an analyte, gave a clear binding response to anti-TNF agents, such as etanercept, infliximab and adalimumab, immobilized on sensorchip. The binding response of the analyte cells increased in a concentration-dependent manner and was competitively reduced by adding soluble TNF receptors to the analyte cell suspension. Treatment of analyte cells with free anti-TNF agent before injection reduced the binding response between the analyte cells and immobilized-etanercept on sensorchip, and the inhibitory effect of free anti-TNF agent was concordant with the affinity of anti-TNF agent for soluble TNF-α. These findings indicate that the SPR response arises from specific binding between anti-TNF agent and its target on cell membrane.
实验方法
使用机型:Biacore T200
实验类型:定性检测,竞争实验
样品类型:配体:蛋白anti-TNF;分析物:Jurkat 细胞
实验芯片:S系列CM5芯片
配体固定方式:标签捕获
运行缓冲液:HBS-EP+
再生试剂:10mM Glycine-HCl, pH 3.5
产品信息
| 类别 | 货号 | 产品名称 | 规格 |
|---|---|---|---|
| 主机 | Biacore T200 | ||
| 芯片 | 29104988 | S系列CM5芯片 | 1片 |
| 芯片 | BR100530 | S系列CM5芯片 | 3片 |
| 芯片 | 29149603 | S系列CM5芯片 | 10片 |
| 芯片 | 29127555 | S系列Protein A芯片 | 1片 |
| 芯片 | 29127556 | S系列Protein A芯片 | 3片 |
| 运行缓冲液 | BR100669 | HBS-EP+ (10×) | 1 × 1000 mL |
| 运行缓冲液 | BR100826 | HBS-EP+ (10×) | 4 × 50 mL |
Reference
Ogura T, Tanaka Y, Toyoda H. Whole cell-based surface plasmon resonance measurement to assess binding of anti-TNF agents to transmembrane target. Anal Biochem. 2016 Sep 1;508:73-7. doi: 10.1016/j.ab.2016.06.021. Epub 2016 Jun 24. PMID: 27349512.