Rapid detection of salmonella enteritidis and escherichia coli using surface plasmon resonance biosensor

作者: C. DEBROY et al.
发表年份:2006年
发表期刊:Journal of Food Process Engineering
作者单位:宾夕法尼亚州立大学等


文章摘要

The Biacore surface plasmon resonance (SPR)-based biosensor was used to detect Salmonella enterica serovar Enteritidis and Escherichia coli in spiked skim milk using specific antibodies. The optical SPR biosensor registers changes in the refractive index during binding of the analyte (bacteria) to the ligand (antibody) on the gold-coated sensor chip surface. Immobilized protein A-captured specific antibodies were used to detect the bacteria in less than an hour. The reaction was found to be specific, and the limits of detection of the assay were found to be 25 cfu/mL for E. coli and 23 cfu/mL for Salmonella. Because the assays were shown to be sensitive, reproducible and very specific, the SPR-based Biacore biosensor has the potential for near real-time, yet specific, detection and presumptive identification of bacteria in food and water.

实验方法
使用机型:Biacore 2000
实验类型:定性检测,筛选实验
样品类型:配体: E. coli or Salmonella Enteritidis的抗体蛋白;分析物:细菌
实验芯片:C系列CM5芯片
配体固定方式:标签捕获
运行缓冲液:HBS-EP+
再生条件:10mM NaOH

产品信息

类别 货号 产品名称 规格
主机 Biacore 2000
芯片 BR100399 C系列CM5芯片 1片
芯片 BR100012 C系列CM5芯片 3片
芯片 29149604 C系列CM5芯片 10片
芯片 29127557 C系列Protein A芯片 1片
芯片 29127558 C系列Protein A芯片 3片
运行缓冲液 BR100669 HBS-EP+ (10×) 1 × 1000 mL
运行缓冲液 BR100826 HBS-EP+ (10×) 4 × 50 mL
试剂 BR100358 50mM 氢氧化钠 1 × 100 mL

Reference
Rapid detection of salmonella enteritidis and escherichia coli using surface plasmon resonance biosensor