作者:王硕等
发表年份:2014年
发表期刊:Journal of Agricultural and Food Chemistry
作者单位:天津科技大学
地区:天津市
文章摘要
A sensitive and stable surface plasmon resonance (SPR) immunosensor based on the inhibition format was developed and validated for detecting melamine (MEL) in milk products and pet foods. The sensitivity and the limit of detection (LOD) of the proposed method for MEL were 2.32 × 10-2and 1.4 × 10-3μg/mL, respectively. The immunosensor was highly specific to MEL, which displayed only low cross-reactivity (CR) (<0.01%) for cyanuric acid, cyanuric chloride, and atrazine. The assay was validated for the detection of MEL in full-cream milk, skim milk powder, infant formula, dog food, and cat food. Most of the recovery results ranged between 76 and 115%. The sensitivities of the assay in each type of sample were 2.57 × 10-2μg/ mL, 2.32 × 10-2μg/kg, 2.51 × 10-2μg/kg, 2.66 × 10-2μg/kg, and 2.68 × 10-2μg/kg, respectively, which were much lower than the maximum residue levels (MRLs) of MEL.
实验方法
使用机型:Biacore X100
实验类型:浓度测定
样品类型:配体:MEL−BSA conjugates蛋白;分析物:melamine/三聚氰胺,奶制品/宠物食品
实验芯片:C系列CM5芯片
配体固定方式:氨基偶联
运行缓冲液:HBS
再生条件:0.1 mol/L HCl
产品信息
| 类别 | 货号 | 产品名称 | 规格 |
|---|---|---|---|
| 主机 | BR110073 | Biacore X100 | |
| 芯片 | BR100399 | C系列CM5芯片 | 1片 |
| 芯片 | BR100012 | C系列CM5芯片 | 3片 |
| 芯片 | 29149604 | C系列CM5芯片 | 10片 |
| 试剂盒 | BR100050 | 氨基偶联试剂盒 | |
| 试剂 | BR100350 | 10mM醋酸钠(PH 4.5) | 1 × 50 mL |
| 运行缓冲液 | BR100670 | HBS-N (10×) | 1 × 1000 mL |
| 运行缓冲液 | BR100828 | HBS-N (10×) | 4 × 50 mL |
Reference
Lu Y, Xia Y, Pan M, Wang X, Wang S. Development of a surface plasmon resonance immunosensor for detecting melamine in milk products and pet foods. J Agric Food Chem. 2014 Dec 24;62(51):12471-6. doi: 10.1021/jf504055g. Epub 2014 Dec 15. PMID: 25473730.