Development of a Surface Plasmon Resonance Immunosensor for Detecting Melamine in Milk Products and Pet Foods

作者:王硕等
发表年份:2014年
发表期刊:Journal of Agricultural and Food Chemistry
作者单位:天津科技大学
地区:天津市


文章摘要

A sensitive and stable surface plasmon resonance (SPR) immunosensor based on the inhibition format was developed and validated for detecting melamine (MEL) in milk products and pet foods. The sensitivity and the limit of detection (LOD) of the proposed method for MEL were 2.32 × 10-2and 1.4 × 10-3μg/mL, respectively. The immunosensor was highly specific to MEL, which displayed only low cross-reactivity (CR) (<0.01%) for cyanuric acid, cyanuric chloride, and atrazine. The assay was validated for the detection of MEL in full-cream milk, skim milk powder, infant formula, dog food, and cat food. Most of the recovery results ranged between 76 and 115%. The sensitivities of the assay in each type of sample were 2.57 × 10-2μg/ mL, 2.32 × 10-2μg/kg, 2.51 × 10-2μg/kg, 2.66 × 10-2μg/kg, and 2.68 × 10-2μg/kg, respectively, which were much lower than the maximum residue levels (MRLs) of MEL.

实验方法
使用机型:Biacore X100
实验类型:浓度测定
样品类型:配体:MEL−BSA conjugates蛋白;分析物:melamine/三聚氰胺,奶制品/宠物食品
实验芯片:C系列CM5芯片
配体固定方式:氨基偶联
运行缓冲液:HBS
再生条件:0.1 mol/L HCl

产品信息

类别 货号 产品名称 规格
主机 BR110073 Biacore X100
芯片 BR100399 C系列CM5芯片 1片
芯片 BR100012 C系列CM5芯片 3片
芯片 29149604 C系列CM5芯片 10片
试剂盒 BR100050 氨基偶联试剂盒
试剂 BR100350 10mM醋酸钠(PH 4.5) 1 × 50 mL
运行缓冲液 BR100670 HBS-N (10×) 1 × 1000 mL
运行缓冲液 BR100828 HBS-N (10×) 4 × 50 mL

Reference